NOTE:
If a red LED is lit on a bay, the battery of the corresponding bay should be
removed from the charger and its contacts checked. Install the battery pack
again. If the red LED illuminates again, remove the battery pack from the
charger and contact ST Technologies
The unit will become warm during normal operation. This does not indicate a
malfunction.
Prevent metallic objects from coming into contact with the metal parts of the DC
plug of the power supply. If this happens, a short may occur and the unit may
be damaged.
Ensure the battery has proper orientation before insertion. Do not attempt to
force a battery into the slot as this may damage the contacts.
5.

CLEANING INSTRUCTIONS

Disconnect the Charger from the power source.
Clean the exterior of the charger with a cloth that has been dampened (not
dripping) using a hospital disinfectant. Follow applicable hospital protocols during
cleaning and dry immediately.
6.

DISPOSAL/RECYCLE

In accordance with the European Waste from Electrical and Electronic Equipment
(WEEE) directive, we encourage our customers to recycle this product whenever
possible. Disposal of this unit must be performed in accordance with the applicable
local environmental regulations.
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/.
Please contact our Repair Department to issue a return authorization to return
product to manufacturer at the end of product life.
7.

TROUBLESHOOTING

SYMPTOM
Battery not
charging
Charger LED
indicator not
working
LIT 232
®
CUDA
Surgical
(English)
POSSIBLE CAUSES/SOLUTIONS
• Ensure charger power cord is plugged in to the AC power outlet.
• Ensure DC connector is plugged into the back of the charger
• Ensure battery is properly seated in the charger bay.
• If repeated attempts to charge the battery have failed, the battery
may be faulty. Replace the battery.
• The charger may have a faulty power cord and/or power brick.
Replace the power cord and/or power brick.
• Ensure charger power cord is plugged in to the AC power outlet.
• Ensure battery is properly seated in the charger bay.
• The charger may have a faulty power cord and/or power brick.
Replace the power cord and/or power brick.
®
for service.
Date of Revision:01/26/18
Rev. -
Page 5 of 31