5.2 Mechanical System

TROUBLE
1. Field of view of one eye
does not match that of the
other
2. Observation is tiring

5.3 Electrical System

TROUBLE
1. The bulb cannot light
2. The bulb burns out
suddenly
3. The illumination is
not bright enough
4. Image flicks
Caution: read instructions (warnings)
carefully
Product complies with European
Directive
Keep away from sunlight
Product code
WEEE disposal
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specific recycling point for electric and elec-
tronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
CAUSE
Interpupillary distance is incorrect Adjust interpupillary distance
The diopter is not proper
The brightness of the illumination
is not proper for eyes
CAUSE
No power supply
The pin of the bulb doesn't insert
properly
The bulb broken
The bulb is not the specified one;
The voltage is too high
The voltage is too low
The bulb is about to burn out
The bulb is not inserted deeply
Symbols
ENGLISH
SOLUTION
Adjust the diopter properly
Adjust the lamp voltage
SOLUTION
Check the power cord connec-
tion
Insert the pin deeply
Replace with a new one
Use the specified bulb; lower the
voltage
Raise the voltage
Replace with a new one
Check its connection
Follow instructions for use
Date of manufacture
Keep in a cool, dry place
Lot number