NOTE: Battery is sold separately. See uline.com
for details.
CAUTION! Read all safety warnings before
installing, making connections or servicing.
Failure to comply with these warnings could
result in personal injury and/or property damage.
Retain all instructions for future reference.
BATTERY INSTALLATION
1.
Disconnect equipment from main power supply
before installation.
2. Separate indicator from base by disconnecting load
cell cable from male connector at back of indicator.
Remove interface option cover. (See Figure 90)
Figure 90
3. Connect seven-pin li-ion battery pack power cable
to main board battery connector. (See Figure 91)
Figure 91
4. Place li-ion battery pack so
slots (1) on bottom right edge
of battery pack align with rib
on bottom of indicator option
area. Slide battery pack to
right (in direction of arrow)
until slots engage with rib.
(See Figures 92-93)
Figure 93
PAGE 31 OF 101
BATTTERY INSTALLATION AND OPERATION
Interface
Option Cover
Load Cell Cable
Male Connector
Cable
Battery
Connector
Figure 92
Slot 1
Arrow
Rib
Boss Post 9
5. Twist li-ion battery pack cable a half turn and
connect seven-pin connector from indicator to
battery pack. Secure battery pack with one screw at
the location marked "Boss Screw 9." (See Figure 94)
Figure 94
6. Put the interface option cover back on the terminal.
BATTERY OPERATION
During operation,
battery icon
indicates charge
level remaining.
The battery icon
Figure 95
is located above
weighing unit on
indicator display.
(See Figure 95)
NOTE: When battery level is below 5% the
indicator will shut down automatically. It takes
about nine hours to fully charge battery.
BATTERY SPECIFICATIONS
Battery Type
Battery Operating Time
Battery Charging Time
Operating Temperature Range
Charging Temperature Range
NOTE: The number of hours the balance can
run on battery depends on many factors.
Backlight and Interface Options will all reduce
the hours before a recharge is needed. For
optimum battery life, turn these features off.
WARNING! Stop charging battery if:
• Charging not completed within specified time.
• Battery becomes abnormally hot.
• There is odor, discoloration or deformation.
• Abnormal conditions are detected during use,
charging or storage.
Li-on Battery
Pack Seven-Pin
Connector
Boss Screw 9
Battery 5-25% remaining
Battery 25-50% remaining
Battery 50-75% remaining
Battery 75-100% remaining
Battery Icon
123
2017 09
19 17 : 31
0.0
>
I1I
g
>
0
X10
I D
Lithium-Ion, 2,500mAh, 7.4V
21 hours with backlight off
9 hours
-4ºF ~ 140º F
32ºF ~ 113º F
0621 IH-8105