R82 Mustang Guia Del Usuario página 27

Ocultar thumbs Ver también para Mustang:
Tabla de contenido

Publicidad

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• Check that the angle mechanism functions
correctly and that the safety lock is working
correctly
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
Monthly
• For products with air-filled tyres. The
function of the anti reverse stop is
dependent upon tire pressure. Check the
markings on the tire and then check the tire
pressure
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
support.R82.org
Disinfection
The product without accessories can be
disinfected with a 70% disinfectant IPA solution.
It is recommended to wipe off any residue
and dirt from the product, using a cloth with
warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Warewashing machine
This product can be washed at:
• 60°C in 10 minutes
...using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories separately.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Castors are moving freely and all castor
locks/brakes are easily activated
• Center bar height and angle is adjustable
and stays in position after adjustment
• Directional lock on each castor
• Nuts and bolts on the product are tightened
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
EN
27

Publicidad

Tabla de contenido
loading

Tabla de contenido