6. Other Functions
6-1 Sleep Function
NOTE
The instrument still consumes small amount of battery power in
the sleep mode. Make sure to set the function selector switch to
the "OFF" position after use.
(1) Sleep Mode
This is a function to prevent the instrument from
being left powered on in order to conserve battery
life. This function causes the instrument to automati-
cally enter the sleep (powered down) mode about
10 minutes after the last switch or button operation.
To exit the sleep mode, turn the function selector
switch back to "OFF", then to any other position, or
press any button.
(2) How to disable the sleep mode
To disable the sleep mode, Power the instrument on
with the data hold switch pressed. "P.OFF" is shown
on the display for about 3 seconds after
the instrument is powered on.
To enable the sleep mode, power the instrument off,
then power it on without pressing the data hold
switch.
6-2 Data Hold Function
This is a function used to freeze the measured
value on the display. Press the data hold switch
to freeze the reading. The reading will be held
regardless of subsequent changes in input 'H' is
shown on the upper left corner of the display while
the instrument is in the data hold mode.
To exit the data hold mode, press the data hold
switch again.
NOTE
If the instrument in the data hold mode enters the sleep mode,
the data hold mode will be cancelled.
7. Battery Replacement
To avoid electric shock hazard, make sure to set the
function selector switch to "OFF" and remove the
test leads from the instrument before trying to
replace the batteries.
All manuals and user guides at all-guides.com
Do not mix new and old batteries
Make sure to install battery in correct polarity as indicated inside
the battery compartment.
When "BAT" is shown on the display, replace the
batteries. Note that when the battery is completely
exhausted, the display blanks without "BAT" shown.
1 Set the function selector switch to the "OFF"
2 Unscrew and remove the battery compartment on
3 Replace the batteries observing correct polarity.
4 Mount and screw the battery compartment cover.
CAUTION
position.
the bottom of the instrument.
Use two new R03 or equivalentbatteries.
Battery Compartment
2XRo3 or 1.5V AAA