4. If your balance is the AHZ-600 model, the display will show "200g".
Put on the weighing pan a calibration weight of 200 g.
5. If your balance is the AHZ-3000 model, the display will show
"1000g". Put on the weighing pan a calibration weight of 1000 g.
6. The display will show automatically "-CAL-".
7. The display´s reading will remain with the value of calibration
weight.
8. The calibration procedure has finished.
PRINTER fUNCTION
By using the RS-232C output, connect the optional printer.
Press the
key to print displayed data.
PRINT CONfIgURATION
Mode 1 "hAnd".
Mode 2 "Auto".
Mode 3 "Contin". The data is continuously transmitted to the printer.
1. Press the
2. The display will show the message "Unit".
By pressing the
transmitted data.
The data will be transmitted automatically when the
weight placed on the pan is > 3d and the display reading
is stable.
key for 3 seconds.
key, the printer prints the
13