Instructions for use
Bionova
MiniPro Auto-reader do not represent photobiological
®
risk and do not generate dangerous optical radiation in any of
its normal operation conditions as per the requirements of IEC
62471 Standard.
Designed under Quality Management System standards ISO
13485:2016/NS-EN ISO 13485:2016.
Product Compatibility and
Incubation programs
Bionova
MiniPro Incubation Programs
®
The following incubation programs are available for your
Bionova
MiniPro. Verify which incubation programs are available
®
for your country checking the Indications for Use section.
Temperature
Time selection
selection
60 ºC
PRO (10 min.)
To select the appropriate incubation program for the Indicator you
are planning to use, follow each Indicator's Instructions for use.
Select the appropriate incubation program before
the incubation of any Indicator.
Bionova
MiniPro Auto-reader compatible indicators
®
We develop new Indicators for extending the features of our
Auto-readers regularly. Please visit www.terragene.com/
compatible or scan the QR code below to check the latest
compatible Indicators developed for your Auto-reader.
For easy identification, each Bionova
a distinctive serial and batch numbers. The number can be
found printed on two labels, one placed at the bottom of the
Auto-reader, and one on its packaging. Please, register your
serial and batch numbers in this manual for future reference:
Serial number:
Batch number:
Incubation Program
PRO at 60°C
Bionova
MiniPro Page 7
®
Serial/Batch Number
Instructions for use
Start-up
1.Place the Bionova
MiniPro Auto-reader on a firm surface,
®
free from vibrations, away from direct sunlight, currents of hot
or cold air, chemical and corrosive or flammable substances.
Do not place the Auto-reader in a way that disconnection of the
power supply's AC plug from mains could be difficult. Leave
a space of at least 10 cm from the Auto-reader to the closest
wall. Do not move the Auto-reader periodically or during its use.
Connect the Auto-reader to a secure and stable electrical mains
outlet socket.
Do not wet or heat the device. If liquid is spilled on the
Auto-reader, disconnect it and follow the instructions
on the Cleaning and maintenance section.
2.Power on your Bionova
MiniPro by connecting the power
®
supply's AC plug to mains and then connect the plug at the
other end of power supply to the rear of the Auto-reader.
Before power on, verify that all incubation positions
are empty.
If the Blue Attention light starts blinking after
powering on the Auto-reader, check that the
printer's door is tightly closed and that the printer
has paper, if not follow the instructions on the
Replacement of the paper roll section.
3.Set the Auto-reader internal clock to correspond to your local
time. See configuration methods in the Time setting mode
section.
MiniPro Auto-reader has
®