yoUr cavitron SySteM iS eQUipped with a water Line fiLter that needS to be evaLUated and
repLaced periodicaLLy.
water QUaLity:
MUnicipaL
Recommended
1 x month
Evaluation and/or
Replacement:
Other
Considerations:
key indicatorS that yoUr cavitron water
fiLter needS to be repLaced:
• Water flow is reduced to the insert and/or handpiece
• If particulates are visible on the filter
• Discoloration of the filter
• Light indicator on your unit is visible
pLeaSe note:
• A 10-pack of replacement filters (part number 90158) is
available from your local DENTSPLY Professional distributor.
• The Cavitron water filter is not intended to be the primary
water filtration for the office. It is meant to protect the
Cavitron system and components.
• If your Cavitron water filter needs to be replaced on a highly
frequent basis, you should consider an overall water filtration
system for your office to minimize damage to your dental
equipment.
• Please refer to the scheduled maintenance guide for
your specific Cavitron system for complete maintenance
recommendations.
to repLace yoUr cavitron water fiLter:
1. Verify that the system is turned OFF.
2. Locate the blue water line on the back of the unit.
3. Disconnect the water supply hose from the dental office
water supply. If a quick disconnect connector is attached to
the end of the hose, relieve the water pressure by pressing
the tip of the connector in an appropriate container and drain
the water.
water filter Maintenance guide
weLL
hard
2 x month
2 x month
Soft
1 x month
2 x month
If using chemical cleaners, the system
should be flushed appropriately to
prevent damage to equipment. Please
refer to the Cavitron Maintenance
Guide for recommended procedure.
4. Grasp the fittings on either side of the filter disk and twist
counterclockwise. Remove the filter section from either side of
the water hose.
5. Install the replacement filter onto the water hose fittings. The
filter should be positioned to match up with the correct hose
fitting.
6. Hand tighten one hose fitting onto filter in a clockwise direction.
Tighten second hose onto filter in clockwise direction.
Reconnect the water supply line, operate unit to bleed the air
and test for leaks.