Restoring Factory Calibration - Refco RESCALE Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

EN
4. While holding
and after a few seconds, the display will show SET CAL and 25.000 kg.
5. Using
or
so that it is equal to the weight of the reference calibration weight.
6. Press
. A number representing the weight of the unloaded platform will
appear on the display. Press
zero.
7. Apply the calibration weight to the platform. A number representing
the applied reference weight will appear on the display. Allow at least
5 seconds for this number to stabilize.
8. Press
. The REFSCALE is now recalibrated to the reference weight.

7.1 RESTORING FACTORY CALIBRATION

If you make a mistake recalibrating the REFSCALE, you can always restore the
factory calibration (and factory defaults) of the scale.
To restore the factory settings perform the following steps:
1. Turn the power to the REFSCALE off.
2. Remove any objects from the platform.
3. While holding
and beep 10 times. The factory settings are now restored.
If the key tone is switched off, no tone sounds, but the calibration
reset has taken place
16 |
REFSCALE Instruction Manual
, press and release
adjust the displayed value (in 1kg increments)
to lock in this value. The display will read
, press and release
. The REFSCALE will turn on,
. The REFSCALE will turn on