WHAT DISPLAY SYMBOLS MEAN

Display
Symbol
MEASUREMENT IN
PROGRESS
LOW BATTERY
ERROR 1
ERROR 2
ERROR 3
NOTE: If you hear an announcement
indicating that a measurement error
has occurred, take the corrective
actions shown above and re-measure.
?
3
Condition/
Cause
Appears during measure-
ment; blinks and beeps
when pulse is detected.
Battery voltage is
too low for monitor
to work properly.
Appears when systolic
and diastolic measure-
ments are within 10
mmHg of each other.
Appears if pressure
value is unstable due
to movement during
reading.
Appears when pressure
value did not increase
during cuff inflation:
1) cuff is not fastened
2) air connector is back-
wards or not connected
properly.
LifeSource Health Line (Toll-Free): 1-888-726-9966
Recommended
Action
Remain still and refrain
from talking; measure-
ment in progress.
Replace all batteries
with new ones.
Check for air leakage; make
sure cuff tube is properly
connected to monitor
and exhaust rate is in
increments of between
2 mmHg. and 5 mmHg
Try measurement again
and remain very still.
Make sure cuff tube is
properly connected to
monitor, cuff is securely
fastened and exhaust rate
is between 2 mmHg and
5 mmHg. Make sure air
connector is inserted
properly into tube
and monitor.