V - STERILIZATION
Tips should be cleaned before sterilization
using a cotton swab and alcohol, wipes or by
immersion in an ultrasonic bath.
Tips can be sterilized by any conventional heat
method.
Conditions
and

sterilization

determined by the equipment:
Equipment: autoclave.
Wet heat sterilization - unwrapped instrument
134°C at 2 bar (200 kPa) for 18 min.
VI - REPLACING TIPS
PM 1, PM 2 and PM 4 tips
These tips must be replaced when the diamond-
coated cutting part becomes smooth and shiny.
PM 3 tip
cycles
are
This tip must be replaced when the cutting part
becomes blunt.