Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to main-
tain a device may compromise the clinical condition or safety of users and/or their caregivers.
EN
EN
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any dam-
aged or worn-out parts
Ensure that all fasteners are in place and fit-
ted correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all wheels are moving freely and
that all wheel locks/brakes are easily acti-
vated
Wipe off any residue and dirt from the prod-
uct using a cloth wrung in warm water and
a mild detergent/soap without chlorine and
dry before use
Disinfect the product. please check the
disinfection section in this User Guide for
more information
Monthly
All nuts and bolt on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substan-
dard or faulty
Warewashing machine
This product can be washed at 60° using a mild
detergent in 15-20 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the acces-
sories seperately.
Disinfection
The product without accesories and cushion can be
disinfected with a 70% disinfectant solution. It is rec-
ommended to wipe off any residue and dirt from the
product, using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry before
disinfection.
ENGLISH
Accessories and spare parts
The products from R82 can be supplied with a vari-
ety of accessories which comply with the needs of
the individual user. Spare parts can be ordered on
request. Find the specific accessories on our website
or contact your local dealer for further information.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in ac-
cordance with the manufacturer's instructions and
demonstrably recorded.
Service Check List:
Check and adjust the following functions as per
their intended use. Repair or replace any damaged
or worn-out parts. Please check the:
Seat plane adjustment. The seat plane is
adjustable and stays in position after adjust-
ment.
If a seat adapter is mounted, check that the
red handle for locking works and that the
safety lock "clicks into position when placing
a seat into the adapter
Wheels are moving freely and that all wheel
locks/brakes are easily activated. Adjust if
required
Quick release wheels are removable/replace-
able
Tire pressure
Spoke guards is without cracks and secured
to the wheel with the clips mounted on each
guard
Before storage: Always deactivate wheel
locks/brakes
Do not place more than one user in the prod-
uct
Anti-tips can be fixed in working and resting
position
Nuts and bolts on the product are tightened
Oil the swing-away parts
Product for cracks or signs for worn-out parts
All labels on the product are intact
Presales information
Please go to support.r82.org