QUICK GUIDE CB011/CB012
G_CB011-CB012_v0003_20211013_EN
Manual
available at:
(22 pages)
https://www.cesva.com/en/support/product/
INCLUDES
POWER AND TURN ON/OFF THE CALIBRATOR
Open the battery compartment
with a screwdriver and insert the
3 batteries respecting the polarity
CHECKING THE DEVICE
Calibrator
CB011 o CB012
1º Insert the device into the calibrator (see figure)
2º Turn on the device*
3º Turn on the calibrator
4º Start measuring with the device*
5º Check that the function value on the screen coincides with the value
If you work with
6º If the difference between the values is within the recommended ranges*, the
verification is favorable. Otherwise, contact the nearest
service.
7º Turn off the calibrator and the device
Maracaibo, 6 08030 Barcelona (Spain)
T. (34) 934 335 240
CB011
2 AA Alkaline
batteries
To turn the calibrator ON /
OFF press
To keep the calibrator on perma-
nently, press
microphones the value corrected is 93,9 dB
info@cesva.com
www.cesva.com
EN
CB012
for few seconds
* Check the device
manual.