man STARTOS PRO EN 28/37
EN
TROUBLESHOOTING
The device does not switch on:
1.
Make sure the plug is firmly fitted to the wall socket;
2.
The thermal protector might have been released. The thermal protector may
be released for the following reasons:
- the device has been operating for a too long time;
- the device was used near heaters or in hot environments;
- the air vent openings are obstructed and do not allow for the motor venti-
lation.
In this case, set the ON/OFF switch to "O" and let it rest for at least 60 minu-
tes.
The device does not nebulize, or it nebulizes very weakly:
1.
Make sure that the ends of the air tube are fitted tightly onto the main unit
and onto the nebulizer;
2.
Check whether the nebulizer is empty from medication or full of an excessive
medication quantity (MAX. 15 ml);
3.
Check whether the nebulizer nozzle is obstructed;
4.
Check whether the air conduction cone is inserted (pic. E).
Should the device not start working properly again, address to an
authorized technical service centre or to your dealer.
MAINTENANCE AND REPAIRS
In case of failure, address to your dealer. Never open the device. There are
no user-serviceable parts within. The compressor requires no lubrication or main-
tenance.
34
16-07-2008
15:32
Pagina 34