Descargar Imprimir esta página

Processing Instructions - Dentatus Profin PDX Manual De Instrucciones

Ocultar thumbs Ver también para Profin PDX:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 33

Processing Instructions

Profin PDX
WARNINGS
Do not immerse the handpiece into disinfection solutions or into ultrasonic batch. The solutions may contain chemicals that may impede or
freeze the handpiece drive mechanism.
Do not use washer disinfector. Anodized aluminium is not suitable for washer disinfector. The hand piece is not tested for repeated cycles in
washer disinfector.
The handpiece shall NOT be disassembled.
Limitations on
Sterilize the handpiece for a maximum of 250 cycles. We recommend that you send the handpiece to the manufacturer for service after 250
processing
uses/cycles, or every 12 months (whichever comes first).
INSTRUCTIONS
Initial treatment at the
Instruction
point of use
1. Flush instrument on unit before removal from tubing.
2. Thereafter remove the tip.
Preparation before
Instruction
cleaning
Ensure the tip is removed from the handpiece.
If heavily soiled, clean first with soaked cloth or pre-soaked wipes.
The maximum period of time between start of use and start of cleaning process shall not exceed 2 hours.
Cleaning: Automated
Equipment
2)
Automated machine for dental handpieces.
Detergent
In accordance with manufacturer's specification for the use of the automated machine.
Instructions
In accordance with the manufacturer's specification for the use of the automated machine.
If soil is visible after cleaning with automated machine, perform manual cleaning.
Cleaning: Manual
External
Equipment
Non-abrasive, medium hard brush.
Detergent
Low alkaline.
Water quality
Deionized (DI) <40°C or Reverse osmosis (RO) <40°C.
Instruction
1. Brush off under non-stagnant water.
2. Rinse under non-stagnant water for 15 seconds.
9
1)

Publicidad

loading

Productos relacionados para Dentatus Profin PDX