Test modes during immunity tests:
Battery mode, standby armed/ready.
Charging mode, actuator not running.
EM immunity standards
Electrostatic discharge
EN
IEC61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity
wireless
equipment
IEC 61000-4-3
Electrical fast transients /
bursts
IEC 61000-4-4
Surges, Line-to-line
IEC 61000-4-5
Conducted
induced by RF fields
IEC 61000-4-6
Rated
power
magnetic fields
IEC 61000-4-8
Voltage
interruptions
variations
IEC 61000-4-11
Variation of power frequency
IEC 61000-4-28
IEC 60601-1-2:2014 test levels
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
fields
form
RF
IEC 60601-1-2:2014 Table 9
communication
± 1 kV, ±2 kV
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
230 VAC
disturbance
6 Vrms
0.15 MHz – 80 MHz
80% AM at 1 kHz
frequency
30 A/m
50 Hz
dips,
short
0 % U
and
voltage
45°, 90°, 135°, 180°, 225°,
270° and 315° phase angles
0 % U
70 % U
0 % U
230 VAC
N/A
for 0,5 cycle at 0°,
T
for 1 cycle
T
for 25 cycles
T
for 250 cycles
T
Additional test levels
± 6 kV contact
20 V/m
800 MHz – 2.5 GHz
80 % AM at 1 kHz
N/A
± 1 kV, ± 2 kV
5 kHz repetition frequency
± 0,5 kV, ± 1 kV
100 VAC and 240 VAC
N/A
30 A/m
50 Hz and 60 Hz
0 % U
for 1 cycle
T
40 % U
for 10 cycles
T
70 % U
for 25 cycles
T
100 VAC and 240 VAC
49 Hz and 51 Hz
100 VAC and 240 VAC, 50 Hz
59 Hz and 61 Hz
100 VAC and 240 VAC, 60 Hz