Calibration
Calibration by Entry of Cell Constant
Input of cell constant with simultaneous display of uncorrected
conductivity value and temperature
Display
10
Action
Press cal key, enter code 1100.
Select with key,
edit number with key,
proceed with enter.
Ready for calibration
Enter cell constant of connected
sensor:
Select using,
enter number using
A change in the cell constant also
changes the conductivity value.
Press enter to confirm cell
constant.
Remark
Device is in the Hold
mode.
If an invalid code is
entered, the device
returns to measuring
mode.
Display (2 sec)
The lower display
shows the measured
conductivity value.
(When there has
not been an entry
for 6 sec, the lower
display alternately
shows the conduc-
tivity and tempera-
ture value.)