DTF PURENEB EXTRANEB+ Manual Del Usuario página 4

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 6
EN
1. Instructions for use
For a good hygiene, before the first use, sterilize the nebulizer in boiling water (20 minutes). Connect the tube to
the "Pressure" outlet of the device and to the inlet (1) of the nebulizer. If connected to the oxygen supply of the
hospital, adjust the flow to the value 6L/min.
Attach the mouth piece (preferred), or the mask, to the aerosol outlet (6). Open the upper part of the body of the
jet nebulizer unscrewing it by ¼ revolution. Check that the jet-pipe (4) is correctly inserted on the bottom part (3).
Pour the medicine in the tank (2) without overfilling the mark "FULL" (6 mL). Place back the upper part of the body
of the nebulizer screwing it by ¼ revolution. Put the device ON. A light mist of aerosol must come out of the mouth
piece or the mask. If it is not the case, check that the jets (3) or (4) are not clogged and that the upper jet (4) is well
positioned down on the bottom part (3). Do not cover the venture hole (5).
See instructions for use of the device.
2. Maintenance
After each inhalation session, clean the jets (3) and (4) by using the device with water. Then, unplug the tube
coming from the compressor. Wash the nebulizer with hot water and let dry. Do not use any metal wire to clean
the jets (3) and (4).
3. Sterilization
Dismantle the body of the nebulizer according to the picture.
Nebulizer, mouth piece, mask and tube can be sterilized in boiling water (20 min).
4. European Standard EN 13544-1
Maximum volume filling: 6 mL
Minimum volume filling: 1 mL
Performances for a 2 mL volume:
Aerosol produced (a)
Flow rate of aerosol produced (b)
MMAD (c)
Particle-size distribution curves shown in the appliance's instructions for use.
(a) Medicine volume inhaled by the patient during one nebulization treatment.
(b) Medicine volume inhaled by the patient during 1 minute of nebulization.
(c) Mass Median Aerodynamic Diameter (size of the aerosols particle).
For hygienic purpose, a brand new EXTRANEB+ nebulizer must be given to each patient.
Nebulizer is compatible with AERODJINN, AERODJINN+ and AEROLIGHT compressors by DTF MEDICAL.
Expected shelf life of accessory: 1 year.
EXTRANEB+ nebulizer
With AERODJINN+ compressor
0.45 mL NaF 1%
0.13 mL/mn
4.5 µm
With AEROLIGHT compressor
0.43 mL NaF 1%
0.09 mL/mn
4.2 µm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Nl9mxbr-gbNl9mxbp-gbNl9mxr-gbNl9mxp-gbNl9mxb-gb

Tabla de contenido