Pro-care 2 / English
7. Maintenance
7.1 General
1. Check main power cord and plug if there is abrasions or excessive wear.
2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are stubbed to-
gether correctly.
3. Check airflow from the quick connector. The airflow should alternate between each connector every
half-cycle time if it's in alternating mode.
4. Check the air hoses if there is kink or breaks. For replacement, please contact our local dealers.
7.2 Fuse Replacement
7.3 Air Filter Replacement
1. Disconnect the plug form mains power when a
blown fuse is suspected.
2. Remove the cover of the fuse holder by means of a
small screwdriver.
3. Insert a new fuse of the correct rating in, and re-
place the cover of the fuse holder back. The fuse
should be rated as T1AL/250V type and VDE ap-
proved.
1. Open the air filter cap located at the back of
pump.
2. Replace with a clean filter and put the cover lid
back. The filter is reusable and can be washed
gently with a mild detergent and water. Dry the
filter before use.
3. Check
and
replace
environment is dirty. Suggested to do it every
three month.
10
air
filter
regularly
if