Troubleshooting Module Installation
1. Power off the oscilloscope.
2. Remove the module and examine the contacts for damage or
contamination. Clean contacts if necessary using a 75%
isopropyl alcohol solution.
3. Reinstall the module into the oscilloscope.
4. Power on the oscilloscope and perform Checking Module
Installation on page 6.
5. Power off the oscilloscope and move the module to another
slot.
6. Power on the oscilloscope and perform Checking Module
Installation on page 6.
DPO4000 Application Module Installation
a. If the Application Module Detected display lists the
application module, cleaning the contacts fixed the
problem.
b. If the Application Module Detected display does not list
the application module, go to step 5.
a. If the Application Module Detected display lists the
module, you have a problem with one of the module
slots. Contact your Tektronix service center.
b. If the oscilloscope does not list the module after you try
the module in every slot, contact your Tektronix service
center.