7 USE IN "LIMIT" MODE
Entering tolerances. Acceptance range
Processing and printing of measurements.Histogram.
NEW DATA
MEMO
U.LT
8,20
L.LT
7.80
N°
VALUE
1
7,96
2
8,03
3
8,09
4
7,93
5
8,01
6
8,06
7
7,96
CANCEL
7
7,98
8
7,88
9
7,94
10
8,11
21
8,88
22
7,97
23
7,64
STATISTICS
N
23
MAX
8,32
MIN
7,71
R
0,61
X
7,9948 M
∆
0,0974 M
s
0,0982 M
-3∆
7,7026 M
+3∆
8,2870 M
L.LT
7,80
U.LT
8,20
-NG
1
+NG
2
P
5,00
Cp
0,684
Cpk
0,666
Cm
0,513
Cmk
0,499
1 - Start-up of the printer and the measuring
instrument
2 - "MODE" switch in "LIMIT" position
M
3 - Press the "START" button
M
All of the data is deleted and the print out of
M
M
M
M
M
4 - Position the measuring instrument on the upper
tolerance value and press "DATA".
M
The tolerance is printed as well as the L.LT.
M
M
5 - Position the measuring instrument on the lower
tolerance value and press "DATA".
M
The tolerance is printed as well as the
MU
6 - Press "DATA" to enter the values.
M
7 - In case of an error, press the "CANCEL"
MU
Measurement above the upper tolerance limit.
M
Measurement below the lower tolerance limit.
ML
STATISTICAL CALCULATION
8 - Press the "STATISTICS" button
The measurement total is printed as well as
the statistical processing of the data.
M
M
M
Number of measurements
Maximum value
Minimum value
Acceptance range
Mean
M
Standard deviation
M
Standard deviation of the population
Number of measurements out of tolerance
Percentage of defective parts
%
Process capability
9 - If the "FEED" button is pressed, the paper is fed
forward without printing.
NEW DATA, MEMO and U.LT appears.
The upper tolerance limit may be entered.
"N°" and "VALUE" .
button before entering other measurements.
The value is deleted.