Sani-tip 122250N Manual Del Usuario página 5

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
Sterilization:
Use a steam autoclave. Place bagged adapter and thumb nut into the steam autoclave, paper side up
when using a paper/plastic pouch. After warm-up is completed, operate at a sterilizing temperature and
pressure of 134ºC/273ºF to 137ºC/279ºF and pressure of 216kPa/31psi for 3 to 12 minutes.
Alternate method: Place non-bagged adapter and thumb nut into the steam autoclave. After warm-up is
completed, operate at a sterilizing temperature and pressure of 134ºC/273ºF to 137ºC/279ºF and pres-
sure of 216kPa /31psi for 3 to 4 minutes. Components should be used immediately.
Drying:
To dry, use the drying cycle of the autoclave. Set cycle for 20-30 minutes.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
• Check for distortion, damage and wear. Discard damaged, worn or corroded components.
Storage:
To maintain sterility, instruments should remain bagged until ready to use.
MANUFACTURER CONTACT
In the United States, contact DENTSPLY Professional Customer Service Technical Support at 800-989-
8826. For areas outside the United States, contact your local DENTSPLY Representative.
TROUBLESHOOTING
PROBLEM: Water leaking, wet air, loose tips and/or loose adapter, indicates either:
a.) Center hole of Sani-Tip disposable tip is not fully seated over water spindle inside new adapter.
b.) Correct Sani-Tip Conversion Kit is not properly installed, and/or missing parts; thumb nut has not
been fully tightened, or incorrect tips are being used.
c.) Adapter is not fully mounted on syringe.
SOLUTION:
a.) Hold adapter to prevent rotation and loosen thumb nut before interchanging Sani-Tip disposable
tips. Insert new tip and push into adapter as far as possible. While holding adapter and pushing in
on the tip, twist tip 1/2 turn clockwise to create a watertight seal over internal tapered water spindle.
Tighten thumb nut.
b.) Make sure correct Sani-Tip
and all O-rings. Verify tips are clear plastic and manufactured by DENTSPLY Professional for the
Sani-Tip
System.
®
c.) Firmly push adapter as far as possible into syringe. Adapter will audibly snap and will fit snugly over
syringe opening when correctly installed.
Conversion Kit is properly installed. Confirm presence of plastic cone
®
5

Publicidad

Tabla de contenido
loading

Productos relacionados para Sani-tip 122250N

Tabla de contenido