Descargar Imprimir esta página

R82 Crocodile Guia Del Usuario página 30

Ocultar thumbs Ver también para Crocodile:

Publicidad

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to main-
tain a device may compromise the clinical condition or safety of users and/or their caregivers.
EN
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any dam-
aged or worn-out parts
Ensure that all fasteners are in place and fit-
ted correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
Wipe off any residue and dirt from the prod-
uct using a cloth with warm water and a
mild detergent/soap and dry before use
Disinfect the product. please check the dis-
infection section in this User Guide for more
information
Monthly
All nuts and bolt on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substan-
dard or faulty
ENGLISH
Disinfection
The product can be disinfected with a 70% IPA disin-
fectant solution. It is recommended to wipe off any
residue and dirt from the product using a cloth wrung
in warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Warewashing machine
This product can be washed at 60° using a mild
detergent for 15-20 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Remove all
accessories before washing the main product.
Clean the accessories separately. Please allow the
product to dry before reuse.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in ac-
cordance with the manufacturer's instructions and
demonstrably recorded.
30
support.R82.org

Publicidad

loading