All manuals and user guides at all-guides.com
Icon
Error
(blinks)
ERR 03
ERR 11
ERR 12
ERR 13
ERR 21
ERR 22
ERR 23
ERR 33
* if output has been set to Alarm = 22mA during configuration
Problem
Possible causes
Temperature probe
Open or short circuit
Temperature range exceeded
Current output 1
Current below 0 (3.8) mA
Current output 1
Current above 20.5 mA
Current output 1
Current span too small / too large
Current output 2
Current below 0 (3.8) mA
Current output 2
Current above 20.5 mA
Current output 2
Current span too small / too large
Sensocheck:
Wrong or defective sensor / polarization
effects at the sensor / cable too long or
defective / plug defective
Temperature outside conversion tables
(TC, conc, SAL)
Error Messages
x
x
x
x
x
x
x
x
Sensoface
active
Sensoface
active
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
13