Carrier Function and Description
CryoPod
Carrier (Item No. BCS-514) is a portable liquid nitrogen (LN2) based system that
is designed to provide over four hours of < -150°C temperature stability for the handling
and transport of cryogenic biospecimens. The system's battery powered electronic module
displays cryogenic chamber temperature, date, time, instrument identification, alarm(s)
and log status. An accessory kit is included for manual LN2 filling. The cryogenic chamber
consists of an anodized metal basket that is seated on the LN2 absorbent pad assembly for
maximum thermal conductivity. When fully charged with LN2 and loaded with one cryobox
or 1-2 small bag cassettes, the CryoPod
built-in handle and compact footprint allows the system to be easily hand-carried over short
distances such as around a laboratory or campus.

Instrument Overview

Power
Set
(A)
(B)
Step 1: Instrument Assembly and Set Up
• Remove the cover from the electronic module using the built-in clip and insert 3 "AA"
batteries into the battery compartment, noting proper orientation. Replace cover.
CAUTION: Batteries could explode or leak if installed improperly, exposing the user
to burns and other risks.
• Turn on the unit by pressing the POWER button(A). Once the LCD display completes its
start-up sequence, 8 operating parameters will need to be confirmed. These steps will
need to be carried out on a new unit, when parameters are updated, and after battery
changes.
• Hold SET button (B) down for 3 seconds to enter Set Mode. "SET" will display on the
LCD. To toggle between each parameter, press SET button (B) again.
Parameter
Description
Alarm 1 (Warning)
To ensure
Program
uncompromised sample
Temperature
handling, two alarm
settings are available.
Program Alarm 1 to
desired warning set
point, which will trigger
when the chamber
warms up and reaches
this temperature.
Alarm 1 (Warning)
The alarm can be
Set On/Off
enabled or disabled.
I
2
English
All manuals and user guides at all-guides.com
LCD screen
Instrument controls
Electronic module
Down
Up
(C)
(D)
Steps
Adjust
temperature by
1°C increments
using DOWN/UP
buttons.
Default: -155°C
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On
Carrier weighs less than 15 pounds (6.8 kg). A
Buttons
DOWN (C)
UP (D)
DOWN (C)
UP (D)
Lid
Handle
Finger grip
LCD screen
Audio
Resolution
Single beep
Add LN2 as needed.
every 30
Alarm will
seconds.
automatically
Pressing the
reset when unit
Up button
temperature is at
can silence
least 5°C colder than
alarm.
warning set point.
Alarms will
Note LCD display
sound for
will have red back-
20 minutes
light, and blinking
before
"Alarm" indicator if
stopping.
triggered.