EN-12
4.2
Sample measurement
4.2.1
pH measurement
Place the electrode in the sample, stir 5 seconds then wait 30-60 seconds.
Press button-
during measurement.
When the reading is stable, press button-
value with the temperature is displayed on the screen, you can record the measurement
result. If you want to do another measurement, press button-
4.2.2
mV measurement
Press button-
mode.
Follow the same procedure as for pH measurement to perform a mV measurement.
4.3
Temperature measurement
For better accuracy, we recommend using a temperature electrode or pH electrode with a
temperature sensor built in.
If a temperature electrode is used, ATC (Auto Temperature Compensation) and the sample
temperature are displayed.
If the meter does not detect a temperature electrode, it automatically switches to the Manual
Temperature Compensation mode and MTC appears. MTC temperature should be set (3.1).
Note: STARTER 2100 accepts NTC 30 kΩ temperature sensor.
to start the pH measurement,
to switch between pH measurement mode and mV measurement
STARTER 2100 Bench pH Meters
appears on the display.
to lock the reading,
is blinking
disappears, the pH
.