LaserLiner AC-Tracer Manual Del Usuario página 9

Ocultar thumbs Ver también para AC-Tracer:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 32
– Observe the safety precautions of local and national authorities relating to the correct
use of the device and any prescribed safety equipment (e.g. electrician's safety gloves).
– Hold the device by the grip sections only. Do not touch the test prods during
measurement.
– Only connect the completely prepared device (transmitter with test leads plugged in)
to a power source. First disconnect the electrical circuit from the power supply and
only switch on again after wiring work has been completed. Lock the master switch
to prevent it being inadvertently switched on.
– Do not operate the transmitter (the device) continuously, but only while actually
capturing images.
– Once the capture is complete, disconnect the transmitter (the device) and the leads
from the circuit.
– Do not work alone in the vicinity of hazardous electrical installations and only under
the guidance of a qualified electrician.
– The measuring device must not be used as a substitute for a two-pole zero
potential test.
– Please be aware that even if a visual signal is not displayed, voltage may still be
present. Differences in the design of the connection socket or the nature of the
insulation (thickness and type) can affect functionality. Voltage cannot be detected
behind panels and metal covers.
– The sender introduces the measuring voltage into the lines to be tested. This may
impair or damage sensitive electronic equipment (e.g. network cards). Therefore,
please make sure that any lines to be tested which belong to sensitive electronic
equipment are isolated before measurement.
– Only use the original adapter.
– Do not use the device in environments containing explosive gases or vapour.
– Protect the device from contamination and damage and store in a dry place.
Additional information on use
Observe the technical safety regulations for working in the vicinity of electrical systems,
especially: 1. Safely isolating from power supply, 2. Securing to prevent system being
switched on again, 3. Checking zero potential, two-pole, 4. Earthing and short-circuiting,
5. Securing and covering adjacent live components.
Safety instructions
Dealing with electromagnetic radiation and electromagnetic disruptions
– The measuring device complies with safety and electromagnetic compatibility
regulations and limit values in accordance with Directive 2014/35/EU (Low Voltage
Directive / LVD) and 2014/30/EU (Electromagnetic Compatibility Directive / EMC).
AC-Tracer
09
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido