LaserLiner ThermoSpot Plus Manual Del Usuario página 9

Ocultar thumbs Ver también para ThermoSpot Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
5
MAX display
The MAX display determines the maximum measured
value within the context of a continuous measurement
(keep trigger button pressed). As soon as you release
the trigger button and press it again to start a new
measurement, the MAX value will be reset and the
instrument will start the process of recording the maxi-
mum measured value all over again.
6
Setting the emission coefficient
A built-in sensor head detects the material/surface-specific infrared rays emitted by every object. The level of
these emissions is determined by the material's emission coefficient (0.01 to 1.00). After switching on for the
first time, the device is preset to an emissivity coefficient of 0.95, which is suitable for most organic materials
as well as plastics, ceramics, wood, rubber and stone. For materials with different emissivity coefficients,
please refer to the table under point 6.
Unknown emission coefficient:
Apply masking tape or black matt paint to the surface of the area whose temperature you wish to measure.
Wait until the tape/paint has heated up. The temperature of the surface can then be measured with an
emission coefficient of 0.95.
!
The emissivity coefficient last selected is set after the device is switched on. Check the emissivity
coefficient setting before each measurement.
...
E 0,01...1,00
ThermoSpot Plus
Laserstrahlung! Nicht in den
Strahl blicken. Laserklasse 2
< 1 mW · 650 nm · EN 60825-1:2014/AC:2017
ThermoSpot Plus
Rev21W11
Button pressed briefly: Value +0.01
Keep button pressed:
Value runs through 0.01 ... 1.00
1x
09
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido