Changing Fuses; Technical Service - B.Braun Aesculap Acculan 3Ti Instrucciones De Manejo

Ocultar thumbs Ver también para Aesculap Acculan 3Ti:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 46
Aesculap Power Systems
Acculan® 3Ti charger GA677

7.1 Changing fuses

Fatal injury from electric shock!
Open components can hold live-threat-
ening mains voltage.
DANGER
Pull the mains plug before replacing
the fuse sets.
Mandatory fuse set:
2 x IEC 127 - T 1.25 A breaking capacity H (1 500 A)
Pull the mains plug before replacing the fuse sets.
Use a small screwdriver to release the clip on the
fuse holder.
Pull out the fuse holder.
Replace both fuses.
Reinsert the fuse holder in such a way that it audi-
bly snaps into place.
Note
If the fuses blow frequently, the product is defective
and must be repaired, see chapter Technical Service.
12
8.

Technical Service

For service, maintenance or repairs, contact your
national B. Braun/Aesculap agency.
Modifications carried out on medical technical equip-
ment may result in loss of guarantee/warranty rights
and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 7461 95-2700
Fax:
+49 7461 16-2887
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood, MO 63042
Aesculap Repair Hotline
Phone:
+1 800 214-3392
Fax:
+1 314 895-4420
Other service addresses can be obtained from the
address indicated above.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ga677

Tabla de contenido