5. Commissioning, operation / 6. Faults
Immunity against radiated fields (DIN EN 61000-4-3)
Under certain conditions, increased indication errors of a maximum of
5 % may occur in the presence of strong electromagnetic fields with a
frequency range of between 80 and 90 MHz. These errors are reversible
EN
and disappear as soon as the source of interference is removed. Do not
install the instrument near sources of strong electromagnetic fields, e.g.
radio transmitters.

6. Faults

WARNING!
Physical injuries and damage to property and the
environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene,
flammable or toxic substances), harmful media (e.g. corrosive,
toxic, carcinogenic, radioactive), and also with refrigeration
plants and compressors, there is a danger of physical injuries
and damage to property and the environment.
Should a failure occur, aggressive media with extremely high
temperature and under high pressure or vacuum may be
present at the instrument.
CAUTION!
Physical injuries and damage to property and the
environment
If faults cannot be eliminated by means of the listed measures,
the instrument must be taken out of operation immediately.
For contact details, see chapter 1 "General information" or the
back page of the operating instructions.
12
For these media, in addition to all standard regulations,
the appropriate existing codes or regulations must also be
followed.
Contact the manufacturer.
If a return is needed, please follow the instructions given in
chapter 8.2 "Return".
WIKA operating instructions model TF-LCD