Treatment - Regular use
2
1
1) Connect the Hose from the
Pressure Chamber to the Control
Unit.
2) Plug in external power supply to
the Control Unit before plugging it
into the wall socket.
7) After the treatment, turn off the
device by pressing the on/off button.
E2 Major Air Leakage
E3 Low Air Volume
Prolonged leakage, treatment is
Not enough air to function (Low air
not being delivered. Solution:
volume)
Check the Seal, Hose with Filter,
Solution: check for obstruction, e.g.
if the Air vents are still in place
kinked hose.
and correct if necessary. Restart
Restart the device by pressing On/
the device by pressing On/Off
Off Button.
Button.
3) Gently pull the Seal over
the outside of the Pressure
Chamber.
8) Roll down the seal onto
the outside of the Pressure
Chamber.
E4 Data transfer error
Data not transferred, e.g. due to taking
out USB Memory Stick too early or
software error
Solution: Switch off the device and try
inserting the USB Memory Stick into
the Control Unit again.
4) Place foot centrally
5) Pull the seal onto
on top of positioner in
the leg to make the
Pressure Chamber,
Pressure Chamber
ensuring air is around
airtight.
the entire leg.
9) Allow evacuation
10)To quickly remove
of the air from the
the air, press down on
Padding by holding
the Padding using
down the release
your leg or other
valve button.
hand.
Timecard | When remaining time
is 25 hours or less, an hourglass
and the time left will show on
display
Solution: Contact your clinician to
load more time with Timecard.
6) Start the
treatment by
switching on the
device by
pressing the on/
off button.
Note:
If the problem persists,
contact clinician or local
distributor. More
information is available
in the manual.