Math Operations; Null - DATATEC EDU34450A Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Math Operations

The table below shows the various math operations that can be used with each measurement function:
Math Operation
DCV

Null

Scaling
Statistics
Limits
– Only one math operation can be enabled at a time.
– When the secondary display is enabled, selecting a math operation will turn off the secondary display.
– Changing the range and resolution is allowed for all math operations.
– You can edit the reference/offset/limit values used for the Null, Scaling, and Limits math operations.
– For remote operation, refer to the CALCulate Subsystem in the EDU34450A Programming Guide.
Null
When making null measurements, also called relative, each reading is the difference between a stored null offset
value and the input signal. One possible application is to increase the accuracy of a 2-wire resistance measurement
by nulling the test lead resistance. Nulling the leads is also particularly important prior to making capacitance
measurements. The formula used for calculating null measurements is:
Null Measurement Display = Reading – Null Offset Value
– The null offset value is adjustable. You can set it to any value between 0 and ±120% of the highest range, for the
present function.
– For resistance measurements, the instrument will read a non-zero value even when the two test leads are in direct
contact because of the resistance of these leads. Short the test leads and press [Null] once the displayed value is
stable to zero-adjust the display.
– For DC voltage measurements, the thermal effect will influence the accuracy. Short the test leads and press [Null]
once the displayed value is stable to zero-adjust the display.
Front Panel Operations
Press [Null] to enable the Null operation. The reading is captured as the null offset value.
68
ACV
DCI
ACI
-
-
Measurement Function
2RES
4RES
FREQ
-
-
-
DIODE
CONT
TEMP
-
-
-
-
-
-
-
-
Keysight EDU34450A User's Guide
CAP
-

Publicidad

Tabla de contenido