Manufacturer's EMC guidelines
Measurement of
interference emissions
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
Magnetic field at supply
frequency (50/60 Hz) in
accordance with
IEC 61000-4-8
IEC 61000-4-11
48
IEC 60601-1 test
Compliance level
level
Not applicable
Not applicable
Not applicable
30 A/m
30 A/m
Not applicable
Electromagnetic environment
guidelines
Magnet fields at mains frequency should
match the typical values found in
business and hospital environments.