Limit the measuring time between 1000 and 1200 A : 15 minutes on, followed by 30 minutes off.
With the C100 model, which is not protected by an output limiter, do not clamp a conductor before
connecting the clamp to the corresponding measuring device. Likewise, do not disconnect the clamp
from the measuring appliance when the clamp is still attached to the cable.
Before connecting the clamp to the multimeter, check that the multimeter has an appropriate calibre.
Open the jaws and clamp the cable through which the current you wish to measure is running. Roughly
centre the cable in the jaws. Follow the direction of the arrow, if so required by the application in question.
To read the measurement, apply the appropriate reading coefficient (See 4.2 "Input/output ratio").
NB : for multi-calibre models, select the ratio which provides the best resolution and precision.
4.1. CONDITIONS DE RÉFÉRENCE
Temperature
Humidity
Conductor centred in jaws
Sinusoidal current
Distorsion factor
Direct current
Continuous magnetic field
Alternating magnetic field
Proximity of external conductors
Measuring device impedance
- C100 / C102 / C103
- C112 / 113
- C122 / C148
- C106 / C107 / C116 / C117
- C160
- C173
4.2. SPECIFICATIONS AND REFERENCES FOR ORDERING
Model
Nominal measuring scope
Alternating current output
C100
100 mA ... 1000 A
C102
C112
1 mA ... 1000 A
C103
100 mA ... 1000 A
C113
1 mA ... 1000 A
C122
1 A ... 1000 A
C148
1 A ... 1000 A
3. USE
4. CHARCATERISTICS
: +20... +26°C
: 20... 75% HR
: 48... 65 Hz.
: < 1%
: no
: earth field (< 40 A/m).
: no
: no current
: ≤ 5 W
: ≤ 1 W
: ≤ 0,2 W
: ≥ 1 MW and ≤ 100 pF
: ≥ 1 MW and ≤ 47 pF
: ≥ 10 MW and ≤ 100 pF
Input/Output ratio
1 mA~/1 A
1 mA~/1 A
5 mA~/1 A~
5 mA~/1 A~
10 mA~/1 A~
20 mA~/1 A~
Connection
Sockets
Sockets
Sockets
Sockets
11
Output protected
against voltage surges
Leads
max. peak
No
30 V