Technical Parameters - Prexiso XL2 Manual Del Usuario

Ocultar thumbs Ver también para XL2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
1.4.3 Vertical accuracy check
As reference, use a plumb-bob and attach it as close as
possible to an approx. 3 m high wall.
Position the instrument at a distance of approx. 1.5 m
from the wall at an elevation of approx. 1.5 m.
Direct the instrument at the wall and switch on. Rotate
the instrument and align it with the bottom of the plumb
line. Now read off the maximum deviation of the laser
line from the top of the plumb line. If the difference does
not exceed 3 mm your Prexiso XL2 is within spec.
8
All manuals and user guides at all-guides.com
max.
3 mm
1.5 m
2

Technical parameters

Functionalities:
Accuracy:
Working range:
Sensitivity range:
Fan angles of horizontal
and vertical directions:
Battery:
Working hours:
Working temperature
range:
Storage temperature
range:
Laser classification:
Dimensions:
Weight:
1 horizontal line plus 1
vertical line
±2 mm/5 m
10 m
auto leveling ±5
greater than 130
3 pieces of AA size
batteries
more than 16 hours (both
lines on)
14°F to 104°F
(-10°C to 40°C)
-4°F to 158°F
(-20°C to 70°C)
Class 2 laser
109*63*87 mm
298 g (excluding batteries)
783114a en