INTRODUCTION

The Console Drive controls the speed of Masterflex
®
L/S
®
pump heads to provide flow rates from 0.10 to 3400 mL/min
or 0.001 to 54 gallons/hr.
Mount up to 2 Masterflex
®
L/S
®
pump heads and all Masterflex-compatible pump heads.

CONTROL/DISPLAY FUNCTIONS

Press keys to activate function. Use INC/DEC (
,
) arrow keys to correct/change a flashing display.
Press any key to enter new values.
A
B
C
D
E
F
Figure 1
L
K
J
I
H
G
A. DOWN ARROW (DECREMENT)—Decrease value of a flashing display.
B. UP ARROW (INCREMENT)—Increase value of a flashing display.
C. FLOW CONTROL—Set flow rate for selected tubing size. To change flow rate, press
or
arrows.
(If pump is running, its speed will change with new settings.)
D. START/STOP—Start/Stop motor.
E. PRIME—Run pump at full speed to fill or clear lines.
F. TOTAL VOLUME—Display cumulative dispensed volume.
G. DIRECTION—To change motor direction.
H. MODE SELECT—INT for internal control; mA for remote current control; V for remote voltage control.
I. UNITS—Select between metric (milliliters and liters) and English (U.S. gallons) displayed units.
J. DISPENSE/COPY—Select dispensed volume, copy volume, or the dispense/copy interval.
K. CAL CONTROL—Refine built-in calibration using a measured volume.
L. TUBING SELECT—Select tubing size.