HEINE mini 3000 Clip Lamp Instrucciones De Uso página 7

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
Setting up
The setting up of the product is laid out in the product overview.
HEINE mini 3000 battery / charging handle (1)
To setting up the mini 3000 handle, unscrew the bottom insert (6) from the handle shell (5) and
insert two alkaline batteries (AA/LR6) or the mini 2Z rechargeable battery with the positive contact
first. Then screw the bottom insert (6) back on.
Charging the mini 3000 charging handle
Setup and operation of the HEINE chargers are described in a separate instruction for use.
Operation
To turn the instrument on, push the switch down; to turn it off, push it back up.
When the clip is clipped onto a breast pocket the handle turns off automatically.
Hygienic reprocessing
The instruction is available:
- at www.heine.com
- in a paper version which you can request from the address listed
Maintenance
Changing the light source
Remove the clip lamp from the handle and pull the lamp out by the collar.
Changing the batteries
Unscrew the bottom insert and shake out the old batteries/rechargeable battery. Insert new
batteries/rechargeable battery with the plus contact facing the handle head.
Service
The device has no components serviceable by the end-user.
General Notes
The warranty for the entire product is invalidated if non-genuine HEINE products or non-original
parts are used and if repairs or modifications are made to the device by persons not authorized
by HEINE. For more information, please visit www.heine.com.
The expected life cycle, when the device is normal used and the warning and safety information
as well as the maintenance instructions are observed, is up to 7 years.
Beyond this period, the product may continue to be used if it is in a safe and good condition.
Do not use alkali manganese batteries containing lithium as battery life will be reduced.
General Warnings
Check the correct operation of the device before use! Do not use the device if there are visible
signs of damage.
Do not use the device in fire- or explosive risk area (e.g. oxygen saturated or anaesthetic
environments)
This product is not allowed to enter or be used in areas with strong magnetic fields e.g. MRI
scanners.
Do not modify the device.
Use only original HEINE parts, spare parts, accessories and power sources.
Repairs shall only be carried out by qualified persons.
The light in this instrument may be harmful. The risk of damage to the eye increases with the
amount of time for which the light is shone into it. An exposure time with this instrument at the
maximum intensity of > 60 s and a distance of > 100 mm will lead to the guideline value for
harm being exceeded.
The HEINE Ear Light is not suitable for the examination of the eardrum and may not be introduced
too deeply into the acoustic meatus.
MED 0811