Repairs; Decommissioning; Disposal - ProMinent DULCOTEST CLE 3.1-mA Instrucciones De Servicio

Medidor de cloro libre
Ocultar thumbs Ver también para DULCOTEST CLE 3.1-mA:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

10 Repairs

The sensor may only be repaired at the factory. Please return in its original
packaging. Prior to return, please carry out instructions as described in

11 "Decommissioning".

11 Decommissioning
Decommissioning sensor: observe all safety guidelines as given in section 5
"Assembly"!
S Disconnect sensor from power supply (see 6 "Installation")
S Depressurise the in-line probe housing
S Loosen the locking screw
S Slowly remove the sensor from the in-line probe housing
S Unscrew the membrane cap over a wash basin or similar and empty
S Rinse the membrane and electrodes with clean water and allow to dry (do
not allow dust to settle on electrodes or membrane)
S Screw on a fresh membrane cap loosely to protect the electrodes
S Fit the membrane cap cover to protect the membrane cap

12 Disposal

Electrolyte
The electrolyte can be disposed of with in-house waste.
Sensor
ProMinent
®
Repairs / Decommissioning / Disposal
IMPORTANT

Electronic waste is classified as special waste!

Observe currently applicable local directives!
31

Publicidad

Tabla de contenido
loading

Tabla de contenido