Biosystems MultiPro con numero de parte 54-49-
103.
1. Verificar que el instrumento esta apagado. Si no
esta apagado, presione el boton de MODO por
tres segundos hasta que aparezca en pantalla
"Release Button".
2. Enchufar la fuente de poder. La luz roja etiquetada
Power cada vez que el cargador este conectado a
la energia.
3. Inserte el MultiPro en el Puerto de carga con la
parte plana hacia abajo y la pantalla hacia el
frente. La luz verde etiquetada "Charge" se
encendera cuando la bateria este en carga.
4. Cuando la bateria este totalmente cargada, la luz
verde "Charge" se apagara.
Ver seccion 5.3.4 para solucion de problemas con
baterias.
5.3.4 Cargando con la bomba conectada
Para cargar el MultiPro con la
bomba motorizada conectada To
charge the MultiPro battery with the
pump installed on the instrument, a
spacer is used in the charger base
to make up the difference between
the size of the MultiPro and the size
of the pump, which is slightly
smaller.
The spacer should be
oriented with the gap towards the
front of the charger, and with the
guide pins facing down towards the
charger as shown in the illustration
at right.
Once the spacer has been inserted,
the MultiPro with pump may be
charged according to the instruction
given in section 5.3.3 above.
5.3.5 Battery troubleshooting
If the green "Charge" LED on the charger fails to light
when the MultiPro with Li-Ion battery pack is placed in
the charger, remove the instrument from the charger
and press the MODE button to attempt to start the
instrument.
If the battery has been inserted into the charger
without the instrument, return it to the instrument prior
to attempting the restart.
1. If the MultiPro starts and the battery icon shows all
three bars, then the battery is fully charged and
may be used.
In this case, the charger has
recognized that the battery is charged and will not
charge it any further.
2. If the MultiPro fails to turn on, then the battery may
be severely discharged and should be returned to
the charger. The charger will then begin a very
slow recharge in order to protect the battery. The
green "Charge" LED may not be lit during the first
four hours of the slow recharge. If the "Charge"
LED has still not been lit after four hours, the
battery pack or charger is probably damaged.
3. If the MultiPro starts and any battery level other
than full is indicated, then either the battery is
damaged or the charger is damaged.
Biosystems for further instructions.
5.4
Sensor installation
The sensors in the MultiPro are located in a vented
compartment at the bottom of the instrument.
To install a sensor:
1. Turn the MultiPro off.
2. Remove the battery pack as described in section
5.1. This will automatically disconnect power from
the instrument.
3. Remove
the
four
screws
shown and pull
the entire main
board
assembly free
from
the
instrument
housing.
4. Gently remove
the sensor that is to be replaced.
5. Insert the new sensor into the appropriate location
on the sensor interface board.
6. Reinstall the main board assembly that was
removed in step three and reinstall the four
screws.
7. Reattach the battery pack.
8. New sensors must be allowed to stabilize prior to
use according to the following schedule.
detector must be powered off and a functional
battery pack must be installed for the sensor to
stabilize.
Sensor
Oxigeno (O
LEL
CO
H
S
2
Note:
Steps 9 and 10 assume that the sensor
stabilization period has passed.
9. If an Oxigeno sensor was replaced, perform the
Fresh Air/Zero Calibracion as discussed in section
4.2.
10. If a combustible (LEL) or a toxic sensor has been
replaced,
perform
Calibracion
and
discussed in sections 4.2 and 4.3.
Note: Sensor channels in the MultiPro are specific
to the type of sensor that occupies the channel.
When replacing sensors, be sure to replace one
sensor with another of the same type.
22
Stabilization Period
) 1 hour
2
none
15 minutes
15 minutes
both
the
Fresh
the
Span
Calibracion
Call
The
Air/Zero
as