10 Measurement

10.1 Procedure

10
1.
Press the POWER button. All
display segments will appear for
approximately one second
before returning to "0"
2.
The unit will automatically inflate
to the appropriate inflation level
based on the user's pulse
oscillations. Measurement will
then begin.
To interrupt the measure-
ment, simply press any key.
The cuff will deflate immedi-
ately after a key is pressed.
During the measurement, do
not talk or move your arm or
hand muscles.
3.
When the measurement is
completed,systolic, diastolic and
pulse will be shown
simultaneously and be saved
automatically in the memory
system. Up to 90 results can be
saved.
4.
Press the POWER button to turn
off the power. If no button is
pressed, the unit will shut off
automatically in 1 minute.
When a mistake was detected
during measurement the Error
mark appears on the display.