MAINTENANCE
The OX-12 requires periodic maintenance. The unit uses an electrochemical sensor with a 2-3 year life expectancy that can be tested,
calibrated and replaced in the field. The OX-12 replacement sensor is available through your local representative or from Macurco. The
detector's performance should be tested regularly by using gas as detailed in the Testing and Field Calibration sections. All other
maintenance and repair of products manufactured by Macurco are to be performed at the appropriate Macurco manufacturing facility.
Macurco does not sanction any third-party repair facilities.
End-of-Warranty Signal
Two years after the OX-12 is installed the sensor end-of-warranty signal will be activated indicating that the OX-12 sensor has reached
the end of its warranty period. The end-of-warranty signal will cause an error code t200 "Sensor warranty expired". See Error Codes
section. In addition to the end-of-warranty signal the sensor itself has a tamper proof warranty expiration sticker dated 2 years from the
unit manufacture date. See the Macurco Fixed Gas Detection Products Limited Warranty section.
The end-of-warranty signal can be silenced for 48 hours by pressing the "ENTER/TEST" button or by temporarily dropping power to the
unit. The end-of-warranty signal provides the user an opportunity to test and/or calibrate the sensor assuring that it is still performing
within acceptable parameters though the sensor is nearing the end of its 2-3 year expected life. The silence function will continue to be
available for 29 days after the OX-12 initiates the initial end-of-warranty signal. After this 29 day period the OX-12 can no longer be
silenced and the sensor must be replaced or the sensor life reset.
Sensor Replacement
1. Remove power to the unit
2. Remove the Philips screw on the front of the OX-12. Pull the front cover of the unit off.
3. Remove the sensor by pulling it gently from the three pronged socket.
4. Power up the unit. The OX-12 steps through an internal self-test cycle for the first 1 minute that it is powered. During the self-test
cycle the unit will display the Firmware Version number, then count down from 60 to 0 and finally go into normal operation. The
indicator light (LED) will flash green during the self-test cycle. At the end of the 1 minute cycle, the unit will take its first sample of
the air and the indicator light will turn solid green.
5. Let the new sensor stabilize for at least 20 minutes then refer to the FIELD CALIBRATION PROCEDURE section to calibrate the
unit.
6. After the successful calibration is complete, reset the sensor life.
Sensor Life Reset
1. Remove the Philips screw on the front of the OX-12. Pull the front cover of the unit off.
2. To reset the sensor life (rSt), from normal or warm-up mode, press the Next button four times to get to SEn or Sensor Mode.
3. Then press the Enter button to get to "rSt" - Reset Sensor Mode.
17