10. Set up the Threshold level microvolt (µV). This is achieved by
contracting your muscles for 5 seconds and taking a reading
of the peak value. The values on the LCD screen will be moving
up and down. Take approximately 40% of the peak value.
E.g. if the reading moves from 25-30 µV, take 40% of 30 µV
which is 12 µV as the threshold level. This level will be different
for each individual.
11. Set the Threshold level by pressing the + or - button.
12. Set the desired parameters by pressing the SET button. This will
move you through the parameters for each mode (VOL, WRK,
RST, TRS, FAB, DOF, WDE – see page 12 for further details).
Press the + or - buttons to change each parameter.
13. Place the unit on a desk in its stand or hold the unit.
14. Relax so that the microvolts reading is as low as you can manage.
15. To start the program, press the START/ESC button. To end
the program before it is complete, press the START/ESC button
again or turn the unit off.
16. When you have finished, remove and replace the electrodes
onto the clear plastic film and reseal them in the plastic bag.
If you are using a PERIFORM
and let it dry before sealing it in its plastic bag. See cleaning
information on page 19.
17. Store the surface electrodes in a cool place, such as a fridge
to extend their 'in-use' life.
NOTE: If the parameters of the unit cannot be changed when
pressing the SET button, the unit will need to be unlocked. Use the
pin from one of the lead wires and press it into the concealed LOCK
button at the back of the unit
PERITONE Instruction Manual - GB
+ or ANUFORM
®
11
, clean the probe
®