Maintenance; Cleaning; Technical Data; Accessories / Spare Parts - Velp Scientifica DLH Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Over Volt
Voltage surge
The software limits the power supply of the motor to the nominal value of the motor itself (60V).
A high speed setting or a heavy load on the chuck may prompt a higher voltage. In this case the voltage is limited to the
nominal value of the motor (60 Volt) and the number of revs is reduced accordingly. In this case,
corner of the display. If for any reason the power supply of the motor exceeds the maximum threshold value of 70V,
"OVERLOAD" the unit will shut down.
NOTE: to reset the unit remove the load from the chuck and turn the unit off and on again using the mains switch.

4. Maintenance

No routine or extraordinary maintenance is necessary apart from periodically cleaning the unit as described in this manual.
In compliance with the product guarantee law, repairs to our units must be carried out in our factory, unless previously
agreed otherwise with local distributors. The instrument must be transported in its original packaging and any indications
present on the original packaging must be followed (e.g. palletized).

4.1 Cleaning

Disconnect the unit from the power supply and use a cloth dampened with a mild, non-flammable detergent.

5. Technical data

Code Model
Admitted power supply
Max. input / output power
Dimensions (WxHxD)
Weight
Speed range at nominal load
Speed indicator / resolution / accuracy
Max torque stirrer shaft
Max stirring quantity H
O
2
Timer with automatic cut-off
Clamping chuck clamping range
Overload and overtemperature protection
Enviromental ambient temperature
Storage temperature range
Max humidity
Operating mode
Electr. safety level CEI-EN60529
Pollution degree / installation cat. CEI-EN61010-1
Max altitude installation
Fuse diam. 5x20mm

6. Accessories / Spare parts

A00001300
Support rod and base
A00001301
Double clamp
A00001302
Ribbon clamp
A00001304
Stirring shaft with floating blade
A00001305
Stirring shaft with folding blade
A00001306
Stirring shaft with fixed blade
A00001307
Stirring shaft with propeller
A00001308
Stirring shaft with paddle,
6 holes
A00001309
Stirring shaft with turbine
A00001310
Stirring shaft with turbo propeller
A00001311
Stirring shaft with anchor
10005266
Knob 26D blue
10000132
Gasket for overhead stirrer
F201A0157
DLH
80-260Volt, 50/60Hz
190 W / 280 W
80x230x196 mm (3.1x9.0x7.7 in)
3 Kg (6,6 lbs)
50 – 2000rpm
LCD / 1rpm / ±2 rpm
80Ncm
40 litres
From 000':00" to 999':59"
From 1 to 10mm
Automatic cut-off
5 - 40°C (41 - 104°F)
-10... + 60°C (14 - 140°F)
80%
Continuous
IP40
2
4000m (13123ft) asl
2x3A
A00001300
A00001301
A00001305
A00001306
A00001309
6
appears in the top rh
F201A0155 DLS
80-260Volt, 50/60Hz
120 W / 180 W
80x215x196 mm (3.1x8.5x7.7 in)
2,5 Kg (5,5 lbs)
50 – 2000rpm
LCD / 1rpm / ±2 rpm
40Ncm
25 litres
From 000':00" to 999':59"
From 1 to 10mm
Automatic cut-off
5 - 40°C (41 - 104°F)
-10... + 60°C (14 - 140°F)
80%
Continuous
IP40
2
4000m (13123ft) asl
2x3A
A00001302
A00001304
A00001307
A00001308
A00001310
A00001311

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

DlsF201a0157F201a0155

Tabla de contenido