Compatibilidad Electromagnética (CEM)
EMV-Allgemeine Daten
Emission
RF-Emission
EN 55011
RF-Emission
EN 55011
Generation of mains harmonics
following IEC 61000-3-2
Generation of mains harmonics
following IEC 61000-3-3
EMV-Tabelle 202
Susceptibility
ESD
IEC 61000-4-2
Bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage drops etc.
IEC 61000-4-11
H-field
at
IEC 61000-4-8
EMV-Tabelle 204
Susceptibility
Conducted
IEC 61000-4-6
Radiated
IEC 61000-4-3
EMV-Tabelle 206
Output
power
transmitter
W
0,01
0,1
1
10
100
Sujeto a modificaciones técnicas
according
according
IEC 60601-test level
±6kV cd
±8kV ad
±2kV mains
±1kV I/O
±1kV dm
±2kV cm
Reduction to
5% for 10ms/ positive
amplitude
5% for 10ms/ negative
amplitude
40% for 100ms
30% for 500ms
0% for 5000ms
50/60Hz
3A/m
IEC 60601-test level
RF
3Veff
150kHz to 80 MHz
RF
3V/mf
80 MHz to 2,5 GHz
of
Safety distance depending on frequency/m
150kHz to 80MHz
0,12m
0,37m
1,17m
3,69m
11,67m
Übereinstimmung
group 1
Class B
Battery operation, not applicable
Battery operation, not applicable
actual level
±2kV,±4kV ±6kV cd
±8kV ad
Battery operation, not applicable
Battery operation, not applicable
Reduction to
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
no m agnetically sesitive sensors
or parts, not applicable
Actual level
3Veff
3V/m
80MHz to 800MHz
0,12m
0,37m
1,17m
3,69m
11,67m
800MHz to 2,5 GHz
0,24m
0,74m
2,34m
7,38m
23,34m
17/56