Electromagnetic Immunity for all EQUIPMENT and SYSTEMS
The REF 73211 is intended for use in the electromagnetic environment specified below. The customer or user
of the REF 73211 should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
ESD
± 6kV Contact
IEC 61000-4-2
± 8kV Air
EFT
±2kV Mains
IEC 61000-4-4
±1kV I/Os
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Voltage Dips, Short
>95% Dip for 0.5 Cycle
Interruptions, and
Voltage Variations
60% Dip for 5 Cycles
on power supply
input lines.
30% Dip for 25 Cycles
IEC 61000-4-11
>95% Dip for 5
Seconds
Power Frequency
3 A/m
50/60Hz
Magnetic Field
IEC 61000-4-8
Note 1: The REF 73211 and its power accessories are a simple power supply and a lamp. The only EMC test
events that could affect the performance or environment of the system are Harmonics, Flicker, Radiated and
Conducted Emission, Surge Immunity, and Voltage Dips, ESD, Radiated RFI, EFT, Conducted RFI, and Power
Frequency Magnetic Immunity should not affect the performance of this product; at the levels required in
EN60601-1-2:2001.
Compliance
Electromagnetic
Level
Environment - Guidance
Not Applicable,
Floors should be wood, concrete,
See Note 1
or ceramic tile. If floors are
synthetic material, the relative
humidity should be at least 30%.
Not Applicable,
Mains power quality should be
See Note 1
that of a typical commercial or
hospital environment.
±1kV Differential
Mains power quality should be
Not Applicable
that of a typical commercial or
hospital environment.
During each dip in
Mains power quality should be
the mains voltage
that of a typical commercial or
the light will
hospital environment. If the user
flicker. During the
of the REF 73211 and its
5-Second dropout,
powering accessories requires
the light shuts off,
that the light remain on during
but returns when
the power outage, the system
mains power is
must be powered from an
restored. See
uninterruptible power supply or a
guidance at the
battery.
right.
Not Applicable,
Power frequency magnetic fields
See Note 1
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
9-13