Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 104

1.6 TECHNICAL DATA

1.6.1 GENERAL
The device is intended for adult, pediatric
and infant populations.
Technical data
Dimensions
Weight
Internal volume
Filter medium
Filter area
Flow range
Pressure
Leakage
Bacterial (BFE)
efficiency
1
Viral (VFE) efficiency
Quality check
Manufacture
Environment
SERVO DUO GUARD, User's Manual
Length = 110 mm,
maximum diameter =
78 mm
55 g
170 ml
Hydrophobic HEPA
filter (High Efficiency
Particulate Air)
Electrostatic filter
955 cm
2
0-3.3 l/s
Max. 30 kPa
<2 ml/min at 15 kPa
over pressure during 5
seconds
99.9999%
2
99.9999%
ISO 23328-1:2003
Each filter is tested
individually for filter
effectiveness and
leakage.
Hygienically
manufactured and
packed.
Incineration of the
device will produce
carbon dioxide and
water.
Servo Duo Guard Instructions
|
Technical data
Storage and transport
temperature
Shelf life
Material
REFERENCES
1
Nelson Laboratories. Bacterial Filtration
Efficiency Test BFE No 350457, with test
values >99.9999%
2
Nelson Laboratories. Virus Filtration
Efficiency Test VFE No 350467, with test
values >99.9999%
Room temperature.
For short periods (1
week): Ambient
temperature -25ºC to
+60ºC. Relative
humidity < 95%.
5 years from
manufacturing date,
when properly stored in
unbroken package
Thermoplastic housing,
hydrophobic glass fibre
paper
| 1 |
7

Publicidad

Tabla de contenido
loading

Tabla de contenido