- Do not overexert the tip on the surface of the teeth in case of hurting the
teeth and damaging the tip.
- After finishing operating, keep the machine working for 30 seconds
on the water supply condition in order to clean the handpiece and the
scaling tip.

3.2 Endo function

Usage process
a.
Fix the endochuck to the handpiece with the torque wrench.
b.
Unscrew the screw cap of the endochuck.
c.
Put the ultrasonic file into the hole of the endochuck.
d.
Screw down the screw cap with the endo wrench to tight up the
ultrasonic file.
e.
Press option key, turn to endo root canal cleaning function.
f.
Press option key, turn the water and vibration intensity to a suitable
level.
g.
Put the ultrasonic file into the patient's root canal slowly, step on the
foot pedal, then make endo root canal cleaning. During the treatment,
turn up the power gradually according to your needs.
Notes
a.
When fixing the endochuck, it must be screwed down.
b.
The screw on the endochuck must be screwed down.
c.
Don't apply to much pressure when inserting the ultrasonic file into the
root canal.
d.
Don't start up the foot pedal until the ultrasonic file is in the root canal.
e.
The power range is supposed from the 1
f.
Please sterilize the endochuck and handpiece after operating.
3.3 Instruction for main components of detachable
handpiece
- Nipple : the nipple can be removed for easy cleaning.
- Light pipe: light guide, can be autoclaved under high temperature and
pressure.
- LED lamp: illuminates the oral area, detachable and sterilized.
- Handpiece: the main part of the whole handpiece can be autoclaved
under high temperature and pressure.
to the 5
level.
st
th
17