RS232 Interface
INTRODUCTION
This Interface Kit is for use with the following Ohaus products: Scout Pro, Traveler
and Navigator.
INTERFACE INSTALLATION
Install the Interface module on the underside of the balance as shown.
Note: The appearance of your model may be different.
1
. Slide Blank Cover to rear.
3
. Install Interface Module.
SETUP
A)
The Ohaus Interface is preset to communicate using the following settings: 2400 baud, 7 bit, no
parity, no handshake. Determine the RS232 parameters required for the printer or computer that
is to be connected. See computer or printer documentation if assistance is required. If the
parameters do not match, it will be necessary to change either the balance settings, or the
computer / printer settings.
B)
Upon installation of the interface, the balance will recognize the RS232 Interface and add
relevant items to the menu. Configure the balance to the desired RS232 and printing
parameters; refer to the balance User Manual for assistance in using the menus.
PRINT
Stable
On, Off
A-Print
Cont, On.Stbl, On.Acc*, 5sec, 15sec,
30sec, 60sec, off
End
Yes, No
*Note: Print and RS232 menu
selections may vary between
the different balance series.
2
. Remove the Blank Cover.
4
. Slide Module forward.
RS232
RS232
On / off
Baud
600, 1200, 2400, 4800, 9600, 19200
Parity
7-even, 7-odd, 7-none, 8-none
Handshake
None, Xon-Xoff, RTS-CTS
END
Yes, No
EN-1