Controlling the medium temperature
The medium temperature is limited by the set safety temperature
controlled by means of a PID controller and measured by the PT
1000 temperature sensor, and the medium is heated up as quickly
as possible without overshooting to the set temperature.
The PID controller adapts to the different tempering media and
ensures optimum temperature control with minimum tempera-
ture drift and fluctuation.
Note: Please comply with the system requirements together with
the operating instructions and help section included with the
software.
USB interface
The Universal Serial Bus (USB) is a serial bus for connecting the
device to the PC. Equipped with USB devices can be connected to
a PC during operation (hot plugging). Connected devices and their
properties are automatically recognized.
Use the USB interface to update the firmware.
The device is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.
Cleaning
Only use cleaning materials recommended by IKA:
Dirt
Cleaning agent
Dyes
Isopropyl alcohol
Building materials Water containing detergent/isopropyl alcohol
Cosmetics
Water containing detergent/isopropyl alcohol
Food
Water containing detergent
Fuels
Water containing detergent
Other materials
Please consult IKA
Wear protective gloves during cleaning the devices.
Electrical devices may not be placed in the cleansing agent for the
purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for clean-
ing or decontamination, you must ascertain with IKA that this
method does not destroy the device.
6

Interfaces and outputs

Maintenance and cleaning

Remove the device from the
mains before cleaning.
Optimum control is only achieved when the medium is mixed us-
ing a rotating evaporator piston.
- Set the desired medium temperature between room tempera-
ture and 99 °C by turning the rotating/pressing knob (C).
- Start the heating function by pressing the rotating/pressing knob (C).
- The LED heating (G) appears.
- The heating bath is heated up to the set temperature.
- The set and actual temperature referred to the medium are
shown on the display.
USB device drivers
First, download the latest driver for IKA devices with USB inter-
face from:
http://www.ika.com/ika/lws/download/usb-driver.zip
Install the driver by running the setup file. Then connect the IKA
device through the USB data cable to the PC. The data communi-
cation is via a virtual COM port.
USB cable A - B
This cable is used to connect the USB interface to a PC.
USB Micro B
Ordering spare parts
When ordering spare parts, please give:
- Device type,
- Serial number, see rating plate,
- Position number and description of spare part, see
www.ika.com,
- Software version.
Repairs
Please only send devices in for repair that have been cleaned
and are free of materials which might present health haz-
ards.
For this, use the "certificate of compliance" form which you
can obtain from IKA or can download a version for printing from
the IKA website at www.ika.com.
If your appliance requires repair, return it in its original packaging.
Storage packaging is not sufficient when sending the device - also
use appropriate transport packaging.
USB A