Understanding Control Test Results
Range (mmol/L)
Level 1
Level 2
Example
The label on your test strip container shows
the acceptable ranges for both the Level 1
and the Level 2 control solutions. The result
you get should be inside this range. Make
sure you compare the result to the correct
level of control.
All manuals and user guides at all-guides.com
1.7–3.4
14.4–19.5
When the control result is inside the range
on the test strip container, your test strips
and your meter are working properly.
Example
53